5.

As the press run continues, observe the

changes in pH and conductivity. When printing

problems such as plugging or scumming

begin, the dampening solution is probably

contaminated. Record your findings in the press

log book and remix a fresh batch of solution.

6.

If alcohol is used in your fountain solution, it

should be added after the solution has been

mixed to the desired conductivity range. Alcohol

has no conductivity, and it dilutes the solution

conductivity value. To properly monitor a

solution containing alcohol while running, the

reading after dilution should be used as the

"standard”.

7.

Drain and clean your dampening system

weekly.

REPLACEMENT/OPTIONS

These items can be obtained from your local dealer:

Item

Description

M6/PH

Conductivity/pH meter (Ranges:

0-5000 µS; 2-12 pH)

512M5

Conductivity meter (Range: 0-5000 µS)

EP

Conductivity meter for DI water system

testing

(5 ranges: 0-0.5, 5, 50, 500, 5000 µS)

EP11/PH

Conductivity/pH meter for checking

reverse osmosis water treatment

systems

(4 conductivity ranges: 0-10, 100, 1000,

10,000 µS; pH range 2-12)

6P

II

Conductivity, TDS, Resistivity, pH, ORP

& Temperature

CLK

Litho-Kit only (without instrument) for

512M5 or other conductivity model

PLK

Litho-Kit only (without instrument) for

M6/PH or other conductivity/pH model

ULK

Litho-Kit only (without instrument) for

Ultrameter 6P

II instruments

pH 4 Buffer*

pH calibration solution

pH 7 Buffer*

pH calibration solution

pH 10 Buffer* pH calibration solution
442-3000*

3900 microsiemens Conductivity

Standard Solution

KCl-7000*

7000 microsiemens Conductivity

Standard Solution

CSFP

Calibrating Solution Four-Pack (1 ea., 2oz

bottles pH 4, 7, 10 & 442-3000)

PS35

Plastic Syringe (35 CC)

TM

Thermometer

M6/pH

512M5

6P

II

*Available in 2oz., quart and gallon sizes