Automation Components RH Test and Calibration User Manual

Rh test and calibration instructions

Advertising
background image

RH Test and Calibration Dip Switch Settings (See Figure 1 & 2)

Note: Do not adjust these switches unless you are using them to troubleshoot or recalibrate the sensor.

Dipswitch #3 should always be left in the ON position during normal operating conditions. Failure to do so

will not allow the RH transmitter to read the sensor (The output will always remain the same).

0% RH Output - Transmitter always outputs a signal of 4 mA or 0 VDC. Sensor doesn’t affect the transmitter

output. (For Trouble Shooting Purposes Only)

50% RH Output - Transmitter always outputs a signal of 12 mA, 2.5 VDC, or 5 VDC. Sensor doesn’t affect

the transmitter output. (For Trouble Shooting Purposes Only)

100% RH Output - Transmitter always outputs a signal of 20 mA, 5 VDC, or 10 VDC. Sensor doesn’t affect

the transmitter output. (For Trouble Shooting Purposes Only)

Normal Operating Condition - The DIP switch must be set in this position for the RH signal to change, due

to the actual measurement of the Relative Humidity by the humidity sensor.

Increment RH Output - This DIP switch will allow you to calibrate the sensor through the software.

The switch must be toggled from the Off to the On position and then returned to the Off position for an

increase of 0.5% RH. This means that if your humidity has drifted 1% over a certain time period, you

will be able to toggle the Increment RH Output switch (2) times in order to slide the whole curve

upward 1%. Note: This is only a single point calibration, and is not recommended for critical applications.

Please contact the factory before doing any field calibration.

Decrement RH Output - This DIP switch will allow you to calibrate the sensor through the software.

The switch must be toggled from the Off to the On position and then returned to the Off position for a

decrease of 0.5% RH. This means that if your humidity has drifted 1% over a certain time period, you

will be able to toggle the Decrement RH Output switch 2 times in order to slide the whole curve

downward. Note: This is only a single point calibration, and is not recommended for critical applications.

Please contact the factory before doing any field calibration.

Reset to Original Calibration - Both the Increment and the Decrement DIP switches should be turned on

for a minimum of 5 seconds before turning them both off again. This will allow you to reset the transmitter

back to the original factory calibration.

Test and Calibration Switch Settings (SW2)

Figure # 1

50% RH Output

6

5

4

ON

3

2

1

Increment RH Output

4

5

6

3

2

ON

1

Normal Operating Condition

4

5

6

3

2

1

ON

0% RH Output

6

5

4

ON

3

2

1

Reset to Original Calibration

6

5

4

1

ON

2

3

Decrement RH Output

4

5

6

ON

1

2

3

100% RH Output

6

5

ON

1

2

3

4

ACI

ACI

“First in Sensor Quality”

Automation Components Inc.

2305 Pleasant View Rd.

Middleton, WI 53562

www.workaci.com

1-608-831-2585

For more information, recalibration

Instructions or to download

any of these instructions go to

www.workaci.com/instructions.htm

RH Test and Calibration Instructions

Advertising