EASYCAL

User Guide Version 4.0 · Status: June 2008

7

EASYCAL

User Guide

3. Main Menu

3.1 Calibration of Liquid Handling Instruments

Click on the gray “Liquid Handling” button , or press F2 to open the “Entry of measuring results” dialog
screen.

3.1.1 Entry of measuring results

Test operator (required entry)

Enter the name of the test operator who carried out the
test. The name of the test operator is automatically stored
in a list.

The Main Menu allows you to access the individual program
features available in EASYCAL

.

To view a list of the test operators that have already been
stored, click on the arrow to the right of the entry field.
You can choose the name of a test operator from this list
and transfer it into the entry field

Serial Number (required entry)

Each instrument is identified by the serial number found on it. You may also enter your own customised identifica-
tion. If you are testing an instrument for the first time, enter the serial number in the entry field. The number will
be stored automatically in a list. Click on the arrow to the right of the entry field to view this list, and to choose a
number already stored.

Instrument (required entry)

The database contains a list of all BRAND GMBH + CO KG Liquid Handling instruments, and of Eppendorf

®

AG,

Gilson

®

Inc., Rainin Instruments, LLC and Biohit Plc. air-interface pipettes. By clicking on the arrow to the right of

the entry field you can view this list and choose the desired instrument. You can also test other instruments that
are not yet included in the list of instruments. To do this, simply enter the name of the instrument in the entry field.
When you save your data, this entry is automatically transferred to the list. By clicking on the arrow to the right of
the entry field you can view the list and choose an instrument which has been saved.