Chemglass AF-0349 Welch Vacuum Pump User Manual

Page 22

Advertising
background image

22

11.

REPLACE THE LIP SEAL HOUSING
a) It is recommended that you use 3 lockwashers to secure the LIP SEAL HOUSING. Uniform tightness is
important. Repeatedly finger-tight the 3 screws in sequence. Then lastly, turn them very tight.
b) Replace the Woodruff Key in its slot. Use your small finger or a thin tool to hold it in place as you replace
the motor coupling housing. Be sure the key stays in position. If it should fall even a little distance, use a
screwdriver to push it back up in the key slot. Then tighten the set key screw—very tight.

Figure 6.1

12.

REPLACE OTHER COMPONENTS, then reassemble the motor and other components.

Advertising