Specifications, Troubleshooting, Warranty – Chemglass CG-3498 User Manual

Page 4

Advertising
background image

4

SPECIFICATIONS:

Sensor Types -



Supply Voltage - 120VAC or 240VAC +/-10%, 50-60 Hz

Dimensions h

x

w

x

d (in) - 3 x 2½ x 5

Ambient Temperature - 0 to 50°C, 32 to 130°F

Fuse - 1A/250V, 5 x 20 mm, fast acting

TROUBLESHOOTING:



Warranty

Chemglass, Inc. warranties this unit against defects in material and workmanship for a period of two years from the date of sale. If
the unit should malfunction, it must be returned for evaluation. If the unit is determined to have a defect in materials or workman-
ship, then it will be repaired or replaced at no charge. Tampering with the unit or damage resulting from excessive current, heat,
moisture, vibration, corrosive materials, or misuse will void this warranty. Programming changes or reconfigurations are not cov-
ered under warranty.

To return this item under warranty, advise the Chemglass technical service department of the situation and they will provide you
with a return authorization number. The unit will be evaluated and repaired as required in accordance to the warranty.

Please contact the Chemglass technical service department with warranty repair information.

Chemglass technical service department:

Tel: 1-800-843-1794 ext. 2124

Fax: 1-800-922-4361

Email: [email protected]

TYPE

RANGE °C

RANGE °F

COLOR

ACCURACY °C

CG-3498-01 (T)

-200 to 250

-273 to 482

Blue

+/- 1

CG-3498-20 (K)

-50 to 1200

-58 to 2192

Yellow

+/- 1

Problem

Possible Cause

Corrective Action

“inPt” “FAiL” flashes

Temperature sensor is unplugged, corroded, or broken

Ensure Sensor is plugged in. Clean, or replace

No display when switch ON

Blown fuse, no power from source

Replace fuse. Check source circuit