7 Stirring

The “stirring” function is switched on or off via rotating

stirring knob. The speed is set on the knob (100 to 1500

rpm in steps of 10 rpm). When both the heating function

and stirring are switched on and the above operations are

completed, the LCD will shift to the speed value and revert

back to the temperature value in about 5 seconds.

The last speed value set is displayed.

8 Timer

Turn the Function switch knob to time set interface, turn the

parameter set knob to target time, press parameter set knob,

to switch on timer function.

Attention: The heating and stirring function can be

switched on at the same time only when setting the timer

function. Otherwise, only the stirring function can be

switched on.

9 Remote Control

Figure 7

The unit can be controlled from an external PC (using the

third-party software) via the RS232C serial interface fitted

to the unit. Data communication from laboratory instrument

to computer is only possible on demand of the computer.

The functions of the interface lines between laboratory

instrument and automation system are selected from

the specified signals of the EIA-standard RS232C,

corresponding with DIN66020 Part 1. The allotment of

the bushing can be taken from Figure 7.

Transmission method: Asynchronous signal

transmission in start-stop-operation.