Addendum, Eurostar control visc analog input – Chemglass CG-16000 User Manual

Page 5

Advertising
background image

3800 North Mill Road • Vineland, NJ 08360 • USA

Phone: 1-800-843-1794 • Fax: 1-800-922-4361

EUROSTAR CONTROL VISC ANALOG INPUT

RPM INPUT CHANNEL 2:

Only use a use a Chemglass analog Eurostar to Data Logger

communications cable when attaching a EUROSTAR CONTROL
VISC Overhead Stirrer to the Data Logger. Note, that it is required
the parameters be changed as listed. The Scale_L and Scale_U may
require adjustment per Eurostar so that both RPM display’s match.

PARAMETERS

Mode=Scale,
Type=Volt,
Range= 6V,
Span_L=0,
Span_U=2,
Scale_L=50,
Scale_U=2000,
Unit=RPM.