4.) Display will not illuminate
(1) Electrical connection - Please disconnect main power
cord and connect again.
(2) Voltage supply - Please check to voltage.
(3) Check the fuse
(4) Verify voltage on the serial number label of the machine
(5) Defective power switch (check voltage at switch)

III-4. Maintenance:
The platform of the Incu-Shaker Mini should be cleaned
regularly with a soft cloth dampened with water or alcohol.
(Always ensure that the machine has been powered down
prior to cleaning.

In the event of a spillage: Power down the unit, remove the
platform and dry the internal chamber with a soft cloth. DO NOT
attempt to operate the unit until the spillage had been properly
dried.





IV-4. Service and Contact