Gb 8. spare part kit / 9. maintenance and cleaning, Maintenance and cleaning – WIKA HP-2 User Manual

Page 18

Advertising
background image

18

WIKA operating instructions pressure transmitter, model HP-2

11379180.03 09/2013 GB/D/F/E

GB

8. Spare part kit / 9. Maintenance and cleaning

Tighten the exchangeable

process connection using the

torque spanner.

On the torque spanner, set

the tightening torque to

130 Nm.

9. Maintenance and cleaning

9.1 Maintenance

Repairs must only be carried out by the manufacturer.

When a leak is suspected, as a result of hairline cracks in the process connection, as

described in chapter 8 “Spare part kit”, replace it.

For the order number for the spare part kit see chapter 12 “Accessories”

9.2 Cleaning

CAUTION!

Before cleaning, correctly disconnect the pressure transmitter from the

pressure supply, switch it off and disconnect it from the voltage supply.

Clean the instrument with a moist cloth.

Electrical connections must not come into contact with moisture.

Wash or clean the dismounted instrument before returning it, in order to

protect persons and the environment from exposure to residual media.

Residual media in the dismounted pressure transmitter can result in a risk

to persons, the environment and equipment. Take sufficient precautionary

measures.

For information on returning the instrument see chapter 11.2 “Return”.