WIKA Operating instructions/Betriebsanleitung/Mode d'emploi/Instrucciones de servicio IS-2X

Failure

Possible cause

Procedure

Output signal unchanged after change

in pressure

Mechanical overload through over-

pressure

Replace instrument; if failure reoccurs,

consult the manufacturer *)

No output signal

No/incorrect voltage supply or current

spike

Adjust the voltage supply to correspond

with the Operating Instructions *)

Cable break

Check connections and cable

No/False output signal

Incorrectly wired

Follow pin assignment (see Instrument

Label / Operating Instructions)

Abnormal output signal

Span incorrectly adjusted

Use appropriate reference

Abnormal zero point signal

Overload limits exceeded

Ensure permissible overload limits are

observed (see Operating Instructions);

correct the zero point through the

potentiometer *)

Failure

Possible cause

Procedure

Abnormal zero point signal

Diaphragm is damaged, e.g. through

impact, abrasive/agressive media;

corrosion of diaphragm/pressure

connector.

Replace instrument

Signal span dropping off/too small

Diaphragm is damaged, e.g. through

impact, abrasive/agressive media;

corrosion of diaphragm/pressure

connector; transmission fluid missing.

Contact the manufacturer and replace

the instrument

Signal span drops off

Seal/Sealing face damaged/conta-

minated, seal mounted incorrectly,

threads crossed.

Clean the seal/sealing face, possibly

replace the seal.

Signal span too small

Mechanical overload through over-

pressure

Re-calibrate the instrument *)

Signal span erratic

Violent fluctuations in the process

media pressure

Damping; consult with manufacturer

9. Maintenance, accessories / 10. Trouble shooting

GB

Accessories: For details about the accessories (e. g. connectors), please refer to WIKA‘s
price list, WIKA‘s product catalog on CD or or contact our sales department.

10. Trouble shooting

Open pressure connections only after the system is without pressure!

Do not insert any pointed or hard objects into the pressure port for cleaning to prevent
damage to the diaphragm of the pressure connection.

„

Take precautions with regard to remaining media in removed pressure trans-
mitters. Remaining media in the pressure port may be hazardous or

toxic!

„

Remove the pressure transmitter from service and mark it to prevent it from
being used again accidentally, if it becomes damaged or unsafe for opera-
tion.

„

Have repairs performed by the manufacturer only.

10. Trouble shooting

GB

In case of unjustified reclamation we charge the reclamation handling expenses.
*) Make sure that after the setting the unit is working properly. In case the error continues to exist send in the instrument for reparation (or

replace the unit).

Please verify in advance if pressure is being applied (valves/ ball valve etc. open) and if the
right voltage supply and the right type of wiring (2-wire) has been chosen?

If the problem persists, contact our sales department.
USA, Canada: If the problem continues, contact WIKA or an authorized agent for assistance.
If the pressure transmitter must be returned obtain an RMA (return material authorization)
number and shipping instructions from the place of purchase. Be sure to include detailed
information about the problem. Pressure transmitters received by WIKA without a valid RMA
number will not be accepted.

Process material certificate (Contamination declaration for returned goods)
Purge / clean dismounted instruments before returning them in order to protect our employees
and the environment from any hazard caused by adherent remaining media.
Service of instruments can only take place safely when a Product Return Form has been
submitted and fully filled-in. This Return Form contains information on all materials with which
the instrument has come into contact, either through installation, test purposes, or cleaning.
You can find the Product Return Form on our internet site (www.wika.de / www.wika.com).