7. Starting, operation

GB

Load (e.g. display)

Power supply

UB+/Sig+

Positive supply / measurement connection

OV/Sig-

Negative supply / measurement connection

Circular connector

M 12x1,

IP 67

Flying leads with 1.5 m of cable,

conducter cross section up

to max. 0.5 mm ², AWG 20

with end splices, conducter outer

diameter 6.8 mm;

cable screen: grey

IP 67

L-Connector,

DIN EN 175301-803, Form A

for conducter cross section up

to max. 1.5 mm², conducter

outer diameter 6 to 8 mm

IP 65

3-wire

2-wire

brown

green

brown

green

white

(D) Mounting hole

1. Loosen the screw (1).
2. Loosen the cable gland (2).
3. Pull the angle housing (5), with the

terminal block (6) inside, away from the

instrument.

4. Using the head of a small screwdriver in

the mounting hole (D), lever the terminal

block (6) out of the angle housing (5).

In order not to damage the sealing of the angle housing, do not try to push the terminal

block (6) out using the screw hole (1) or the cable gland (2).

5. Ensure that the conductor outer diameter you select is matched to the angle housing’s

cable gland. Slide the cable through the cable gland nut (2), washer (3), gland seal (4) and

angle housing (5).

6. Connect the flying leads to the screw terminals on the terminal block (6) in accordance

with the pin-assignment drawing.

7. Press the terminal block (6) back into the angle housing (5).
8. Tighten the cable gland (2) around the cable. Make sure that the sealing isn’t damaged

and that the cable gland and seals are assembled correctly in order to ensure ingress

protection.

9. Place the flat, square gasket over the connection pins on the top of the instrument

housing.

10. Slide the terminal block (6) onto the connection pins.
11. Secure the angle housing (5) and terminal block (6) to the instrument with the screw (1).

Assembly of L-connector DIN EN 175301-803

(6)

(5)

(1)

(2)

(3)

(4)

Clamping nut,

Male connector,

Case with

pressure connection

Sealing

Female

connector

7. Starting, operation