WIKA Operating instructions DELTA-plus, DELTA-comb and DELTA-switch

4

!

Caution

1



740.01

Illustration for example

st

308

15 07/

009 GB/D/F

GB

1. Safety instructions ... 3. Operating principle

1. Safety instructions

The appropriate national safety regulations (i.e. VDE 0100 /

EN 60 079-14 / EN 837-) must be observed when installing,
commissioning and operating these instruments.

Do not work on gauge while under voltage (applies to Model 70.0/03.100

and 851.0.100)

Serious injuries and/or damage can occur should the appropriate regulations

not be observed

Only appropriately qualified personnel should work on these instruments

2. General

These operating instructions are based on the following information:

EN 837-: Selection and installation recommendations for pressure gauges

Data sheet PM 07.15: Differential pressure gauge with integrated working

pressure gauge su Model 70.01.100

Data sheet PV 7.16: Differential pressure gauge with integrated working

pressure gauge and micro switch

T

Model 70.0.100

Data sheet PV 7.17: Differential pressure switch sv

Model 851.0.100

3. Operating principle

The j and i measuring medium chambers

are separated by an elastic diaphragm.

The differential pressure causes an

axial movement (measured travel) in the

diaphragm against the measuring range

spring. The measured travel is proportional

to the differential pressure and is trans-

mitted, „pressure sealed“ and with minimal

friction, via a connecting rod; for:
su

to the

movement

st

to the

movement

and in addition to the plungers
on the

microswitches

sv

to the plungers

on the

microswitches