WIKA operating instructions differential pressure transmitter, model DPGT40

17

14093267.01 03/2014 GB/D/F/E

6. Commissioning, operation / 7. Options and accessories

6.3 Setting up a voltage supply

The voltage supply is made via a power supply unit or a control unit which

provides the energy limitation.

The power supply for the pressure transmitter must be made via an energy-

limited electrical circuit in accordance with section 9.3 of UL/EN/IEC 61010-1, or

an LPS to UL/EN/IEC 60950-1, or class 2 in accordance with UL1310/UL1585

(NEC or CEC). The power supply must be suitable for operation above 2,000 m

should the pressure transmitter be used at this altitude.

7. Options and accessories

7.1 4-way valve manifold

Isolation of the ⊕ and ⊖ process lines for removing or testing the measuring

instrument without interrupting the running process operation.

Protection of the instrument against

excessive overpressure loading, such as

in pressure tests and undefined operating

conditions (including intermittent shutdown).

Pressure compensation for zero point

checking with running processes, and

avoiding one-sided overpressure loading

during start-up and operation phases (with

opened pressure compensating valve).

Venting the measuring lines with liquid

media and flushing of the measuring lines,

in order to remove contamination.

Vent valve

Shut-off valve

⊖ side

Shut-off valve

⊕ side

Pressure

compensating valve

Specifications for handling

Sequence of operations to start measurement

1. Open the pressure compensating valve (middle valve spindle)

2. Open the shut-off valve for the negative media chamber (⊖, right-hand valve)

and the positive media chamber (⊕, left-hand valve)

3. Close the pressure compensating valve

Sequence of operations to flush/vent the measuring lines

1. Start: Open the shut-off valve for the ⊕ and ⊖ media chamber, open the

pressure compensating valve and vent valve

2. Finish: Close the pressure compensating valve and vent valve