6. Commissioning, operation

( + )

( - )

( + )

( - )

LOOPPWR

CEP 1000

MODE

%ERROR

AUTO

MAX

25%

MIN

Calibration Line

SIM

mA

3) The output signal can be adjusted to the set point by means of the knob. For this

purpose select the decade to be set using the arrow keys.

4) The output signal can also be increased incrementally in steps of 4 mA (25 %) using

the "25%" key.

5) The automatic step function and automatic ramp function allow automatic change of

the output signal.

For further information refer to the section "6.3.1 Automatic operation modes" of the

instruction.

6.2.2 Milliampere simulation

Figure 4 shows the way the CEP1000 is connected if it is supposed to function as a

2-wire transmitter with an external power supply. In this case the circuit is powered via

an external voltage source and the instrument just regulates the current in the circuit.

The regulating procedure of the output signal is the same as for the application as the

source of current in the milliampere range.

Figure 4: Simulation of a sensor with a current output for 4 ... 20 mA

Power supply DC 30 V

max.

The instrument is operated in the simulation mode as a

two-wire sensor.

Instrument with an input

of 4 ... 20 mA

4 ... 20 mA