7. Interface

Command

Answer function

Menu settings
:smm <a | c | m >

Define the operating modes of the instrument

a --> start automatic mode

c --> start control mode

m --> start measure mode

:ssw <1 ... 100>

Step size

1 ... 100 --> step size from operating the up and down buttons in %

:swm <a | c | m | z l v | s> Define the operating modes of the instrument

a --> start automatic mode

c --> start control mode

m --> start measure mode

(only available in CONTROL and MEASURE modes)

z --> ZERO

(only available in CONTROL mode)

l --> LEAK test

l --> back to CTRL mode (like LEAK button)

v --> VENT (vent the entire system)

(only available in AUTOMATIC mode)

:szc <0,1>

Zeroing before entry into control mode

0 --> off

1 --> on

:szi <1 ... 60>

Zeroing interval

1 ... 60 --> time in minutes

7.5 Driver

The current USB interface driver is "Future Technology Devices International Limited"

(FTDI) provided for download on its website for all major operating systems (Windows,

Linux, Mac OS).

Homepage: http://www.ftdichip.com

Download: http://www.ftdichip.com/Drivers/VCP.htm (FTDI Homepage --> Drivers -->

VCPDrivers)

Installation instructions for the various operating systems:

http://www.ftdichip.com/Documents/InstallGuides.htm (FTDI Homepage -->

Documents --> Installation Guides)