Gb 6. commissioning, operation – WIKA BCS10 User Manual

Page 20

Advertising
background image

20

WIKA operating instructions calibration system, model BCS10

14051046.01 10/2012 GB/D/F/E

GB

6. Commissioning, operation

Function locking

The calibration system model BCS10 is delivered from the factory with unlocked access

to all settings and they can be changed.

The access to the adjustable parameters can be limited or completely prohibited to

prevent unauthorized changes of the configuration.

This is achieved by the function locking.

In addition, user defined units can be scaled here.

For configuration of the function locking see chapter 6.7 "Configuration of the function

locking".

6.7 Configuration of the function locking

Scroll to the menu item "FUnC

LOCK

" and press the ENTER key. On the digital indicator

appears the message "0PWRD". Enter the password "101". The password is factory set

and cannot be changed.

The functions that can be locked or unlocked:

- zero point adjustment (ZERO button)

- setting the units

- auto Power-Off Settings

- setting of the damping

- setting of the sampling rate

- tare setting

The command LOC locks the respective function and UnLOC unlocks it again.

Locked functions are not displayed in the configuration menu any more and therefore

cannot be changed.

Setting of a user-defined unit or a user-defined measuring range:

In the menu item SET FACTOR, a multiplier between 0.001 and 100 can be set for

a user-defined measuring range. The defined factor is multiplied by the measured

pressure value in psi. The result is displayed.

Example: 40 psi correspond to 1.000 g of a product in a tank. The weight of the product

should appear on the display.

If 25 is defined as the multiplicator, the pressure of 40 psi corresponds to the value

1.000 (40 x 25). „CUST“ is displayed on the instrument as a technical unit.