Young Sensors Relative Humidity & Temperature Probes Models 41342V User Manual

Young Sensors

Advertising
background image

R. M. YOUNG COMPANY

2801 AeroPark Drive , Traverse City, Michigan 49686 USA

TEL (231) 946-3980 FAX (231) 946-4772

MODEL 41342VC/VF

PLATINUM TEMPERATURE PROBE 0-1V Output

INSTRUCTION SHEET 41342V-90

REV 09-97

INTRODUCTION

The Model 41342VC/VF Platinum Temperature Probe is an accurate
1000 ohm Platinum RTD temperature sensor and low power voltage
interface circuit mounted in a weatherproof junction box. The probe is
available in Celsius or Fahrenheit calibration. Output signal is 0-1 VDC
full scale. The probe is designed for easy installation in YOUNG Multi-
plate and Aspirated Radiation Shields.

INSTALLATION

For accurate measurements, the temperature probe should be installed
in a protective radiation shield. Use of the probe without a radiation
shield may result in large errors due to solar heating. For best
performance, the probe and shield should be placed in a location with
good air circulation clear of large masses (buildings, pavement, solar
panels...), exhaust vents, electrical machinery, motors, water fountains
and sprinklers.

MAINTENANCE

The temperature probe is designed to offer years of service with minimal
maintenance. Temperature calibration should be accurate for the life of
the probe. If necessary, the probe may be periodically checked or
recalibrated using normal bath calibration methods. NIST traceable
calibration is available from YOUNG at nominal cost.

WARRANTY

This product is warranted to be free of defects in materials and construc-
tion for a period of 12 months from date of initial purchase. Liability is
limited to repair or replacement of defective item. A copy of the warranty
policy may be obtained from R. M. Young Company.

CE COMPLIANCE

This product has been tested and shown to comply with European CE
requirements for the EMC Directive. Please note that shielded cable must
be used.

SPECIFICATIONS

Power Required:

8 - 24 VDC, 5 mA

Calibrated measuring range:

-50 to +50°C

-50 to +150°F

Accuracy at 0°C:

±0.3°C (differential measurement)

Time Constant:

42 seconds in 43408 shield.

Sensor type:

1000 Platinum RTD

Output signal:

0-1 VDC

Recommended Cable:

2 pair shielded, 22 AWG (#18723)

Recommended Radiation Shields:

Model 43408P

Gill Aspirated Radiation Shield

Model 41002P

Gill Multi-Plate Radiation Shield

Declaration of Conformity

Application of Council Directives:

89/336/EEC

Standards to which Conformity is Declared:

EN 50082-1 (IEC 801-2, 3, 4)

Manufacturer's Name and Address:

R. M. Young Company
Traverse City, MI, 49686, USA

Importer's Name and Address:

See Shipper or Invoice

Type of Equipment:

Meteorological Instruments

Model Number / Year of Manufacture:

41342 (V, L)/1996

I, the undersigned, hereby declare that the equipment specified
conforms to the above Directives and Standards.

Date / Place: Traverse City, Michigan, USA February 19, 1996

David Poinsett
R & D Manager, R. M. Young Company

Advertising