Next Advance DX User Manual

Page 12

Advertising
background image

O

PERATOR

S

M

ANUAL FOR THE

F

REEDOM

R

OCKER

DX

CHANGING FROM 4-TRAY to 2-

TRAY OPERATION (Freedom

Rocker DX4 and DX4-MD only)

The tubing manifold has a small
knob. Turn the knob clockwise,
push it inwards, then turn it
counterclockwise to lock it in 2-tray
operation. To return to 4-tray
mode, turn the knob clockwise, pull
it outwards, then turn
counterclockwise to lock it in place.

OTHER TRAYS AND BOTTLES

A list of commonly requested trays
and bottles is available on our
website and price list. If these do
not fit your needs, please contact
customer service by phone or at

[email protected]

and we

may be able to provide you with a
custom solution.

TROUBLESHOOTING

In addition to those given below,
there is a thorough list of
troubleshooting tips at
http://www.nextadvance.com/FAQs/
FAQs-Freedom-Rocker.htm.
If nothing happens, all the lights
are off and the display is blank
,
check that the power is properly
connected. If the wall transformer
is plugged in all the way and the
power line is plugged into the
instrument, then either the wall
outlet or the wall transformer is
bad. Plug something else into the
same wall outlet. Does it work? If
so, then the power supply is
probably bad. Contact us for a
replacement.

If liquid is not dispensed
immediately after the pump is
turned on
, the tubing was probably
empty and needed to be primed. If
this is the case, the problem should
not occur as long as the tubing does
not fill up again with air.
If liquid is not dispensed or
aspirated
, the caps are probably
tightly screwed onto the bottles and
air is not able to enter or leave the
bottles and equalize the pressure.
Loosen the caps on the bottles. If
not caught early, this could lead to
tubing popping off a fitting or
leakage at a fitting.

SUPPORT

A thorough list of FAQs is at
http://www.nextadvance.com/FAQs/
FAQs-Freedom-Rocker.htm.
If you cannot find a good answer
there, please contact customer
service by email at
[email protected] or by
telephone.

SPECIFICATIONS

The following values are nominal
values. Actual values may vary
slightly.
Size: 12 in. wide x 13.7 in. deep x
10.5 in. high (DX4); 10 in wide x
13.7 in. deep x 10.5 in. high (DX2).
Weight: 15 lbs.
Dispense Flow Rate: 7.5 mL/s
(model DX2); 12.8 mL/s (model
DX4)
Rocking Speed: 0 to 35 cycles per
minute
Power: 12VDC, 3.7 Amp min.

(DX4); 3.0 Amp min. (DX2)