Then, connect the luer lock fittings
to the aspiration head and the
others ends onto the One>Click™
caps on the bottles.

ASPIRATION HEADS

The aspiration head can be swung
out of the way when accessing the
trays.

The quick release pin (white arrow)
holds the aspiration head to the hinge.

Replacement
To switch aspiration/dispense heads,
first disconnect the tubing from the
head by undoing the luer lock
fittings. Then simply remove the
quick release pin by pulling on the
ring. Next, place the desired
aspiration head on the platform, line
up its transverse hole with that in
the pivot block, and replace the
quick release pin. Finally, reconnect
the luer lock fittings.

Sanitizing
To sanitize the tube set or the
aspiration head, use a standard

sanitizing solution, e.g. 5% bleach
or 30% ethanol, followed by
thorough rinsing with distilled water.
Please note that the tube set and
the aspiration head should not be
autoclaved because the temperature
is too high for some of the plastic
parts.

OTHER TRAYS AND BOTTLES

A list of commonly requested trays
and bottles is available on our price
list. Please contact customer
service by email at
[email protected] or by
telephone to discuss your needs.

TROUBLESHOOTING

In addition to those given below,
there is a thorough list of
troubleshooting tips at
http://www.nextadvance.com/FAQs/
FAQs-Freedom-Rocker.htm.
If nothing happens, all the lights
are off and the display is blank
,
check that the power is properly
connected. If the wall transformer
is plugged in all the way and the
power line is plugged into the
instrument, then either the wall
outlet or the wall transformer is
bad. Plug something else into the
same wall outlet. Does it work? If
so, then the power supply is
probably bad. Contact us for a
replacement.
If liquid is not dispensed
immediately after the pump is
turned on
, the tubing was probably
empty and needed to be primed. If
this is the case, the problem should
not occur as long as the tubing does
not fill up again with air.