Operator’s manual for the pressure injection cell – Next Advance PC77-MAG User Manual

Page 5

Advertising
background image

3

BE SAFE. ALWAYS MAKE SURE THAT THERE IS NO PRESSURE IN THE CELL BEFORE LOOSENING THE

BOLTS, AND THAT THE BOLTS ARE SECURELY FASTENED BEFORE APPLYING PRESSURE TO THE CELL.

USERS MUST ALWAYS WEAR PROTECTIVE APPAREL, INCLUDING EYEWEAR AND MUST BE TRAINED IN

WORKING WITH HIGH PRESSURE.

Operator’s Manual for the Pressure Injection Cell

The sample tube protrudes from the base for easy

insertion and removal.

REPLACING THE TOP

Confirm that the black O-ring is seated in

its groove. Slip the top onto the base, rotate

the top 60 degrees clockwise, and then

tighten the three bolts using a hex wrench.

Use a hex wrench to tighten the three bolts.

INSERTING A CAPILLARY

Loosen the nut covering the white ferrule

using a wrench. Slide the capillary through

the small hole in the ferrule, until it reaches

the bottom of the sample tube. Tighten the

nut, thereby squeezing the ferrule tightly

around the capillary.

To remove the capillary, loosen the nut and

slide the capillary out.

INJECTING A SAMPLE

First point the knob on the 3-way valve

towards the stainless steel tubing connected

to the gas tank. Then, while looking away

from the gas tank gages and with your hand

covering them (in case they explode when the

pressure is turned on) open the gas tank

valve. Adjust the pressure regulator to the

SETTING UP YOUR PRESSURE

INJECTION CELL

Setting up the Pressure Injection Cell

consists of attaching it to 1/8 inch (outside

diameter) thick walled stainless steel tubing

(not supplied with the PC77 Pressure

Injection Cell). First, slide the tubing through

the inlet nut until it hits a stop. Finger tighten

the nut and then tighten it with a wrench for

1-1/4 turns. Connect the other end of the

tubing to a pressurized gas tank. It is now set

up.

USING YOUR PRESSURE

INJECTION CELL

OVERVIEW

Using the pressure injection cell requires

removing the top, inserting your sample,

replacing the top, and then inserting a

capillary. Next, inject the sample using

pressurized gas to force the sample through

the capillary. After the injection, close the gas

tank valve and open the 3-way valve to

release the pressure.

REMOVING THE TOP

With the valves at the pressure source

turned off and the 3-way valve opened to

release any pressure inside the cell, loosen the

three bolts securing the top, 2 turns each.

Rotate the top counterclockwise by 60

degrees so that the hexagonal sides are even.

Then lift off the top.

INSERTING THE SAMPLE

Place the sample tube in the main hole of

the base. The tube should not be capped. If

the cap is hinged, you must cut off the hinge

(with scissors). There is a recess in the top so

the tube can protrude from the base, but a

hinged cap will not fit in the recess.