External adjustment – RADWAG MAC 210 Moisture Analyzer User Manual

Page 29

Advertising
background image

29

10.1.

External adjustment

External adjustment in moisture analyzer MAC series should be carried out

with an external mass standard / weight class F2 or higher.

Start of external adjustment process,

On selecting CAL mode, the moisture analyzer displays a command unLoAd

ordering unloading of the weighing pan (the weighing pan must be empty).

When the weighing pan is empty press Print/Enter key. The moisture analyzer

initiates adjustment procedure. In a moment the display indicates mass of a

standard / weight to be loaded on the weighing pan – on loading the mass

standard / weight press Print/Enter key. On completing adjustment process

the moisture analyzer returns to submenu P1-01 ECAL, (if a moisture analyzer

is connected to a printer or a computer, then it automatically sends a report

from adjustment process to be printed – only if corresponding option has been

enabled in moisture analyzer’s settings).

CAUTION
- Press ESC key to abort adjustment process.
- If during adjustment process, weighing pan of a moisture analyzer is
loaded, then moisture analyzer displays an error message Er 1 Hi.


Return to weighing mode
(see point 7.2 – return to weighing mode).