Zeroing, Weighings in two ranges, Selection of basic weight unit – RADWAG Low profiled scales User Manual

Page 23: Eighings in two ranges, Eight unit, Selection of basic w

Advertising
background image

11.3. Zeroing

To ZERO the scale press:

.

The scale will display zero and following pictograms:

and

.

Zeroing is only possible within the scope of ±2% of full scale. While
zeroing outside the scope of ±2% you will see <Err2>. Zeroing is
possible only in stable state.


Notice:
Zeroing is possible only within the ±2% interval of the maximal range.
If zeroing is performed beyond this range the <Err2> message and
short audible signal will be emitted.



11.4. Weighings in two ranges

Switching between the I range and the II range happens automatically
(exceeding Max of the I range). Weighings in the second range is
signalled by a pictogram in the top left corner of the display.

Then weighings is done with the accuracy of the II range to the moment
of returning to zero (autozero range

) where the scale switches back

to the I range.



11.5. Selection of basic weight unit

This function is used to set weight unit the scale will start with.

Procedure:

• Enter the submenu <P3.Unit> and then:

- 23 -