RADWAG PUE C41H Indicator User Manual

Page 157

Advertising
background image

HOLD B.

NO

CHUTE ENABLE

NO

• Enter the submenu <P8.2 OUTPUT CONF.> and configure outputs

according the table below:

MIN NO

OK NO

MAX NO

STABLE NO

THRESHOLD 1

OUT1

THRESHOLD 2

OUT2

CHUTE OUT3


2. Return to weighing according to ch.11.3.

3. Connect the cable Scale - I/O (see ch. 25) and then install external

buttons START, STOP (terminal inputs) and subsequent devices
to the terminal outputs according to the drawings below:


4. Enter

the

DOSING operating mode according to ch. 24.8.6,

5. Press

to enter thresholds setting:

MIN – mass value over which the bulk dosing changes to the precise

dosing ,

MAX – mass value over which dosing is completed and CHUTE starts.

6. Before you start dosing unload the pen (container, silo) or tare the

scale pressing

.

7. Start the process by pressing the external START button,

pictogram starts to blink, and threshold 1, THR 1 bulk dosing,
is performed.

- 157 -