2 distortion test, Cdb5374 – Cirrus Logic CDB5374 User Manual

Page 12

Advertising
background image

CDB5374

12

DS862DB1

• Once the Setup panel is set, select Configure on the Digital Filter sub-panel.

• After digital filter configuration is complete, click Capture to collect a data record.

• Once the data record is collected, the Analysis panel is automatically displayed.

• Select Noise FFT from the Test Select control to display the calculated noise statistics.

• Verify the noise performance (S/N) is 121 dB or better.

1.4.2

Distortion Test

• Set the controls on the Setup panel to match the picture:

• Once the Setup panel is set, select Configure on the Digital Filter sub-panel.

• After digital filter configuration is complete, click Capture to collect a data record.

• Once the data record is collected, the Analysis panel is automatically displayed.

• Select Signal FFT from the Test Select control to display the calculated signal statistics.

• Verify the distortion performance (S/D) is 108 dB or better.