With the Pump/Drive Assembly removed from the
instrument and the piston/cylinder group removed
(MAY REQUIRE PUMP RECALIBRATION):

b) If the pump is severely frozen then the piston/cylin-
der group must be removed from the pump. Remove
the (2) screws that secure the H435 Cap to the pump
base assembly and carefully remove (twist, tilt & pull)
the piston/cylinder group from the pump base assem-
bly.

c) Soak the piston/cylinder group in warm water for
20 minutes minimum. For quickest results use an
ultrasonic bath. Then try to loosen up the assembly by
hand - gently rotating the piston back and forth until
free.

CAUTION:
Do not pull piston out of cylinder or force the
piston as breakage or seal damage will occur.

d) Reassemble the piston/cylinder into the pump
base assembly (use FMI Instructions H431) page # 8
section # 22 for more details relative to installing the
piston/cylinder assembly.

e) Reinstall H435 Cylinder Cap and apply a suitable
thread lock to the screws (Loctite # 425) and secure
screws (torque to 9 in-lbs).

f) Rotate the assembly to assure it moves smooth
and free. (If not see note #1 below).

g) Check the pump for proper calibration (If out of
specification see note #1 below).

Note#1:
Please return complete pump/drive unit to the fac-
tory for rework and
recalibration

IN-ST431-14

4

CAUTION

͗ŽƌŝĞŶƚ͞,ϰϬϱ͟ůĞƚƚĞƌŝŶŐ

as shown

H435 Cap

Screws(2)

Tighten to

9 in-lbs torque

www.fmipump.com

[email protected]