32

2

Discard the test strip properly.

Note: You may also turn the monitor

off by Pressing and Holding the

button. If you do not turn your

monitor off or pull the test strip out,

the monitor shuts off automatically

after 60 seconds.

Understanding Your Result

Compare the control result to the "Expected Results for

Use with Solartek or MediSense Control Solutions" range printed

on:

The blood glucose test strip instructions for use, for

glucose control solution testing.

If the control result falls within the range:
What It Means
Your monitor and test strips are working correctly.

If the control result

does not fall within the range:

What It Means

Your monitor and test strips may not be working correctly.

What to Do
Repeat the test with a new test strip and ensure the

control solution testing instructions are followed

completely. If the result is still not within the printed

range, contact Customer Care.