2.) Operational Instructions
1.) ON/OFF key:
A.) Press this key to turn on or to turn off scale.
B.) If the scale is ON, press and hold this key to turn the scale OFF
2.) TARE key
A.) Press this key to zero out weight displayed. (This feature can be implemented when adding items
or weight to hook. This will display the weight of new item/weight only, to attain individual
weight.) (ie a weighing container)
3.) MODE key
A.) Press this key to switch between the weighing modes. (Lbs/KG)
4.) LED key
A.) Press this key to activate LED flashlight.
ERROR DISPLAY
If the display shows “EEEEE” the scale has been overloaded. Remove the weight immediately to avoid
permanent load cell damage.
LOW BATTERY
If the display shows “LO” the battery life has expired. Replace with new batteries.
Calibration
This scale has been calibrated at the factory. If you wish to recalibrate your scale please reference the
calibration instructions.
Calibration Instructions
Note: You will need a 20KG calibration weight in order to properly calibrate this scale.
1. Make sure the scale is powered OFF.
2. Press and hold the MODE and ON/OFF/TARE key simultaneously until the display shows the A/D
value (Random Number) then release both keys.
3. Press the ON/OFF/TARE key once, the display will show “CAL” then “20.00”. Apply the weight.
Once the weight is stable press ON/OFF/TARE key again.
4. The display will show “PASS” the “20.00”.
5. Calibration is complete.