2.4 CCD Gain
This adjustment is to match the sensitivity of the CCD to the luminance of the
installation location.
1
Assemble the Adjustment Plate and place it on the cradle.
NOTE !
For the procedure to assemble the Adjustment Plate, refer to “2.1
Assembling the Adjustment Plate” and “2.2 Placing the Adjustment
Plate” in Chapter 7 “2 Appendix”
When the Glass Plate Unit has been attached, open the Glass Plate Unit, and place
the Adjustment Plate on the cradle.