Temperature-Measurement range:

Measurements in the “ear area” : 34~42.2°C (93.2~108°F),

Scan- operation: -22~80°C (-7.6~176°F).

Operation temperature range: 10~40°C (50~104°F), relative

humidity ≤85%

Storage temperature range: -20~50°C (-4~122°F), relative

humidity ≤85%.

Transport temperature should be lower than 70°C, relative

humidity ≤95%.

Atmospheric pressure: 800~1013 hPa

In accordance to ASTM E1965-98, EN12470-5:2003

Clinical Thermometers Part 5: Performance of Infra-red ear

thermometers (with maximum device), IEC/EN60601-1-2
(EMC),

IEC/EN60601-1 Safety Guidlines.

Deviations

With measurements in the ear: +/-0.2°C (0.4°F) within the

range of 35.5~42°C (95.9~107.6°F),

+/-0.3°C (0.5°F) out of this range.

In Scan-mode +/-0.3°C (0.5°F) within the range of

22~42.2°C (71.6~108°F),

+/-4% or +/-2°C (4°F) out of this range—

the higher value is decisive.

Fever Alarm saving locations for ear operation (display of

date/time/operation mode for each measurement).

Time display, °C / °F toggle function.

Battery, 1 Lithium cell (CR2032 1 pc.).

Battery life: around 3,000 readings. Enable normal use: 1 year

Expected Service Life: 4 years

Measurement in the ear, the device will calculate the

corresponding value which would be derived from an oral

measurement.

Battery low warning.

Automatic switch-off of measurement activation.

Enclosure Rating: IP22

Dimensions: 146.5 x 38.2 x 54.7 mm

Weight: 55.5 grams including battery

The device should not submerge into any liquids and

expose it to direct moisture.
This symbol indicates important for persons or the

device!

FT 4919 AEG Ear Thermometer (TH520B)