Before You Begin

1.

Have all the materials you will need ready before you

begin testing. This includes your Ascensia C

ONTOUR

Meter,

the Ascensia C

ONTOUR

Test Strips, and the Ascensia

M

ICROLET

Lancing Device with Ascensia M

ICROLET

Lan-

cets. You may also need Ascensia

®

C

ONTOUR

Control

Solution to run a quality control check.

2.

Read your user guide and all instructional materials pro-

vided in your meter kit before testing.

3.

Let the meter and test strips adjust to the temperature

of the location where you are testing. Whenever the
meter is moved from one location to another, allow 20
minutes for the meter to adjust to the temperature of
that new location before performing a blood glucose
test. The products have been designed to give accurate
results at temperatures between 41°F and 113°F.

4.

It is important not to use the test strips or control

solution if the expiration date printed on the bottle
label and carton has passed or it has been six months
(180 days) since you fi rst opened the bottle. It will help
to write the six month

discard date on the label in the

area provided when you fi rst open the test strips or
control solution.

5.

Wash your hands thoroughly with warm soapy water

and dry them well.

Potential Biohazard

Healthcare professionals or persons using this system on
multiple patients should follow the infection control pro-
cedure approved by their facility. All products or objects
which come in contact with human blood, even after
cleaning, should be handled as if capable of transmitting
viral diseases.