Accuracy

The accuracy of the Ascensia

C

ONTOUR

Blood Glucose Monitor-

ing System has been assessed in a correlation study conducted

by healthcare professionals. In this study, 108 fresh capillary

blood specimens were tested with the Ascensia

C

ONTOUR

Blood

Glucose Monitoring System by 108 untrained people with dia-

betes using three test strip lots. The reference measurement

was done on the YSI Glucose Analyzer. Ascensia

C

ONTOUR

data

compare well with the results from the YSI method.

Regression Statistics (Plasma/Serum Reference):

plasma reference glucose range: 4.2 to 22.0 mmol/L

Precision

A study was conducted with the Ascensia

C

ONTOUR

Blood

Glucose Monitoring System with fi ve heparinized whole

blood specimens with glucose levels from 2.2 to 18.0

mmol/L. Multiple replicates (n=100) were tested using mul-

tiple Ascensia

C

ONTOUR

Blood Glucose Meters and one lot

of Ascensia

M

ICROFILL

Blood Glucose Test Strips. The fol-

lowing precision results were obtained:

Plasma/Serum:

Technical Information:

Technical Information

Lot

y =

Correlation

Coeffi cient

(r)

1

1.02x – 0.01 mmol/L

0.972

2

1.03x – 0.24 mmol/L

0.963

3

0.95x – 0.02 mmol/L

0.974

Mean

SD

Coeffi cient

of

variation

2.1 mmol/L

0.10 mmol/L

4.8%

5.6 mmol/L

0.28 mmol/L

5.1%

6.5 mmol/L

0.24 mmol/L

3.7%

11.4 mmol/L

0.38 mmol/L

3.3%

18.1 mmol/L

0.65 mmol/L

3.6%