Calling up the saved alarm events

>

Menu 01

Setpoint

>

Menu 06

Settings

>

Menu 05

Alarm

>

The last alarm which occurred is shown in the

display.

30 alarm events are saved. The highest figure (A _ _) is the last
alarm which occurred. Select a previous alarm event by pressing
the buttons

and

.

>

A01

Door Open

S hh:mm

DD/MM/YY

Start of the alarm event with time and date.

>

A01

Door Open

E hh:mm

DD/MM/YY

End of the alarm event with time and date.

>

A01

Door Open

Temp Max

- - °C

Highest temperature reached inside the appliance.
Return to the standard indication by pressing the

button.

Note

After this query, the alarm display will be deleted automatically.

Menu 05
Alarm

= list of alarms

The alarm events are saved and can be called up if required.

The following alarm can be rectified immediately:

Door Open
Temp.

- - °C

When the door is open, the above indication and the standard

indication flash alternately. After 240 seconds, the audible warn-

ing signal sounds.
Cancel the audible warning signal by pressing button

. Close

the appliance door.

The following alarm events require the stored items

to be checked:

1.

ALL Power cut
Temp.

- - °C

The above indication and the standard indication flash alternately.
This indication is displayed after a power failure.
Press button

to delete this display.

Call up the length of the power failure and the maximum tempera-

ture reached inside the appliance. See chapter "Calling up the

saved alarm events".

2.

High Temp Room
Temp.

- - °C

The above indication and the standard indication flash alternately.
The audible warning signal sounds.
Cancel the audible warning signal by pressing button

.

This can be caused by:

• the appliance door being left open for a long time, allowing warm

ambient air to enter;

• a long power failure;

• a fault in the appliance.
If a large quantity of warm items is placed in the appliance at once,

this alarm may occur. This does not mean that there is an error.

Call up the length of the temperature rise and the maximum tem-

perature reached inside the appliance. See chapter "Calling up

the saved alarm events".

In both cases, the items must be checked and it must be

decided if they can still be used.

All other alarms which occur must be rectified by a customer

service technician.

Alarm test

To check that the alarm function is working correctly, an alarm

test should be carried out at regular intervals.

5 seconds

>

Alarm Test

The display flashes and the audible warning signal sounds. The

alarm test is terminated automatically after a few seconds.