AM1860/1850 Standard Platinum Resistance Thermometer

User’s Guide

8/10

5 Troubleshooting

5.1 Troubleshooting

If the SPRT functions abnormally, it could be caused by several

possible problem conditions that are described in this section. Try

the solutions recommended and if the problems are still not solved,

contact manufacture for warranty or repair service. Be sure to have

the model number and serial number of your SPRT available.

5.2 Problem Causes and Solutions

a. R0/Rtpw becomes higher significantly. This is likely caused

by mechanical shocks. The SPRT should be annealed to

release the stress of platinum wires and to recover R0/Rtpw

value. Measure R0/Rtpw of the SPRT after annealing to

verify.

b. R0/Rtpw unstable during the measurement. This is likely

caused either by bad connections or sensor coils short. Check

the connections first and if the connections are good, the

SPRT may be damaged.