Model 745 Digital Delay Generator.

November 2013

3- F

RONT

P

ANEL

O

VERVIEW

The Model 745 front panel is composed of several kinds of elements: the user interface which

consists of a keypad and an LCD display that allows the user to program various settings and to

interact with the device in local mode, 4 status LEDs, 5 BNC connectors.

Figure - 1.

Front panel

LCD screen

A 2 line x 20 character module displays parameters for programming operations and status

information.

Keypad

A six touch keypad provides access to the configuration menu:

-

▲▼ are used to select the line/function to display. After a press on the ‘OK’ button,

up/down increments/decrements digit at the current cursor location

-

◄► are used to move the cursor position among digits

- Ok sets Sub-Menu and ‘cursor mode’, or if this mode is already set, it stores the current

value and clear the cursor mode. Transition from Menu item to Sub-Menu item.

- Esc clears the cursor mode without storing any value, gets out from the Sub-Menu item or

run/stop a channel pulse generation.

BNC Connectors

One Trigger Input and 4 independent output channels are available. The output voltage is tunable
from 2 to 5V for each output independently and each one has to be terminated in 50Ω.

The input Trigger BNC connector provides a trigger signal operating up to 50 kHz.

Status LED indicators

One status LED by output channel, each LED blinking at the output pulse frequency.