Troubleshooting – BNC 1703M User Manual

Page 13

Advertising
background image

1703M OPERATING MANUAL

6. WARRANTY

Berkeley Nucleonics Corporation warrants all instruments, including
component parts, to be free from defects in material and workmanship,
under normal use and service for a period of one year. If repairs are
required during the warranty period, contact the factory for component
replacement or shipping instructions. Include serial number of the
instrument. This warranty is void if the unit is repaired or altered by others
than those authorized by Berkeley Nucleonics Corporation.

Warranty does not cover any instrument once the warranty period has
expired; if evidence of opening the instrument by the user is apparent;
or in the event of mechanical damage if the requirements of operation
were not satisfied.

Warranty does not cover battery replacement.

5. TROUBLESHOOTING

Table 2. Actions to take if instrument is not operating properly.

12

Problem

The instrument fails to

exit self check

The LCD displays “OL” on

the upper scale or “999”

on the lower scale

Audible and/or vibration

alarm are not operating

LCD displays the battery

discharge warning sign and
alarm operates

Possible Cause

1. Gamma detector block failure
2. Processing block failure

1. A gamma radiation source is
too close to the unit

2. Gamma detector block failure
3. Processing block failure

1. Alarm is turned OFF
2. Alarm failure

1. Exhausted battery

Solution

Return instrument for repair.
Return instrument for repair.

Remove the source or step
back 1 foot at a time until the
“OL” message goes away
Return instrument for repair.
Return instrument for repair.

Turn the alarm on
Return instrument for repair.

Replace Battery