Technical data – IKA EUROSTAR 200 control P4 User Manual

Page 17

Advertising
background image

17

EUROSTAR 200 control

EUROSTAR 200 P4 control

EUROSTAR station + Wireless Controller (WiCo)
Speed range I (high torque)

Speed range I (high speed)

rpm

0 / 6 – 400

0 / 30 – 2000

0 / 4 – 110

0 / 16– 530

Speed adjustment

Stepless

Speed display

TFT / Wireless Controller

Speed setting accuracy

rpm

± 1

Deviation – speed measurement

Speed < 300 rpm: ± 3 rpm / Speed > 300 rpm: ± 1%

Max. torque at stirrer shaft

Speed range I

Speed range II

Ncm

200

40

660
130

Torque trend measurement

yes

Torque trend display

yes

Deviation – torque measurement

Speed range I

Speed range II

Ncm

± 20

± 6

± 60
± 10

Max. stirring quantity (water)

ltr

100

Max. viscosity

mPas

100000

150000

Intermittent operation

yes

Reversible direction of rotation

no

Port for ext. temperature sensor

yes

Temperature display

yes

Timer function

yes

Permitted on time

%

100

Nominal voltage

VAC

230 ± 10% (EURO) / 115 ± 10% (USA)

Frequency

Hz

50 / 60

Max. input power

W

130

134

Max. output power at stirring shaft

W

84

76

Protection class according to DIN EN 60529

IP 40

Protection class

I

Excess voltage category

II

Contamination level

2

Protection at overload

yes / motor current limitation

Fuse (on mains plate)

A

T 4 A (IKA

®

Ident. No. 2585100)

Ambient temperature

°C

+ 5 to + 40

Ambient humidity (rel.)

%

80

Drive

Brushless motor

Clamping chuck – clamping range

mm

0,5 – 10

Hollow shaft internal diameter

mm

10.3

no

Extension arm (Ø x L)

mm

16 x 220

Housing

Coated aluminium casting and thermoplastic plastic

Dimensions (W x D x H), without extension arm,
with Wireless Controller

mm

91 x 231 x 294

91 x 231 x 379

Weight (with extension arm and clamping chuck)

kg

4.9

5.8

Operation at a terrestrial altitude

m

max. 2000

USB interface

yes

RS 232 interface

yes

Temperature measurement resolution

K

0,1

Measurement range, temperature

°C

- 10 to + 350

Limiting deviation, temperature sensor
PT 1000 DIN EN 60751 class A

K

≤ ± (0.15 + 0.002 x ITI)

Measurement accuracy, temperature

K

± 0.5 + tolerance PT 1000 (DIN EN 60751 class A)

max. communication range (dependent on the building)

m

40 – 150

Dimension (B x T x H) – Wireless Controller

mm

71 x 74 x 151

Weight – Wireless Controller

kg

0.28

USB interface – Wireless Controller

yes

RB 1 Battery pack
Voltage

V

3.7

Battery capacity

mAh

2000

Charging time

h

4.5

Working time

h

15

Battery type

Lithium-polymer

Subject to technical changes!

Technical data