Troubleshooting
Centrifuge 5418 / 5418 R
English (EN)

32

7.2

Error messages

If one of the following error messages appears, proceed as follows:

1. Remove fault (see Remedies).
2. If necessary, repeat centrifugation.

Symptom/message Cause

Remedy

LID ERROR (5418)/
LID (5418 R)

Centrifuge lid cannot be locked.

Try again to close centrifuge lid.

LID ERROR (5418)/
LID (5418 R)

Centrifuge lid cannot be released.

1. Switch the centrifuge off and back on.
2. Press the

open key.


If the error occurs again:

1. Switch off centrifuge.
2. Activate the emergency lid release

(see

Emergency lid release on p. 33).

LID ERROR (5418)/
LID (5418 R)

Centrifuge lid must not be released
during a run.

Wait for rotor to stop.

INT

Mains/power failure during a run.

Check the mains connection.

NO RPM (5418)/
Error

3 (5418 R)

Error in the speed measuring system or
drive overheated.

Leave the device switched on until the
error message disappears (10 s or 6
min).

Error 5
(only for 5418 R)

Prohibited opening of lid or lid switch is
defective during a run.

1. Wait for rotor to stop.
2. Open and close again the lid of the

device.

3. Repeat the run.

Err 6 (5418)/
Error

6 (5418 R)

Drive fault.

Repeat the run.

If this error message appears again,
switch centrifuge off and back on
again after > 20 s.

Error 6
(only for 5418 R)

Drive overheated.

Allow the drive to cool down for at
least 15 min.

Err 7 (5418)/
Error

7 (5418 R)

Major deviation in the speed control.

1. Wait for rotor to stop.
2. Tighten rotor.

Err 8 (5418)/
Error

8 (5418 R)

Drive fault.

1. Wait for rotor to stop.
2. Repeat the run.

Err 9 to 17 (5418)/
Error

9 to

17 (5418

R)

Electronics error.

Switch centrifuge off and back on
again after > 20 s.

Error 18
(only for 5418 R)

Too high temperature deviation from set
value in the rotor chamber.

Check the settings.

Check unhindered air circulation
through the air slots.

Thaw ice or switch off device and
allow it to cool down.