Maintenance

Centrifuge 5418 / 5418 R

English (EN)

27

6

Maintenance

6.1

Prepare cleaning/disinfection

The procedure described in the following chapter applies to the cleaning as well as to the disinfection or
decontamination. The table below describes the steps required on top of this:

6.2

Cleaning/disinfection

Cleaning

Disinfecting/decontamination

1. Use a mild cleaning fluid to clean the accessible

surfaces of the device and the accessories.

2. Carry out the cleaning as described in the

following chapter.

1. Choose the disinfection method which

corresponds to the legal regulations and
guidelines in place for your range of application.
For example, use alcohol (ethanol, isopropanol)
or alcohol-based disinfectants.

2. Carry out the disinfection or decontamination as

described in the following chapter.

3. Then clean the device and the accessories.

If you have any further questions regarding the cleaning and disinfection or decontamination
or regarding the cleaning fluid to be used, contact the Eppendorf AG Application Support. The
contact details are provided on the back of this manual.

DANGER! Electric shock as a result of penetration of liquid.

Switch off the device and disconnect the power plug before starting cleaning or
disinfection work.

Do not allow any liquids to penetrate the inside of the housing.

Do not spray clean/spray disinfect the housing.

Only plug the device back in if it is completely dry, both inside and outside.

NOTICE! Damage from the use of aggressive chemicals.

Do not use any aggressive chemicals on the device or its accessories, such as strong and
weak bases, strong acids, acetone, formaldehyde, halogenated hydrocarbons or phenol.

If the device has been contaminated by aggressive chemicals, immediately clean it by
means of a mild cleaning agent.

NOTICE! Corrosion from aggressive cleaning agents and disinfectants.

Do not use corrosive cleaning agents, aggressive solvents or abrasive polishes.

Do not incubate the accessories in aggressive cleaning agents or disinfectants for a longer
period of time.