4 rotors, 1 rotor a-4-81 (only 5810/5810 r), Operating manual – Eppendorf 5810 R Centrifuge User Manual

Page 13

Advertising
background image

13

Centrifuge 5804/5804 R/5810/5810 R — Operating manual

2.4

Rotors

You can operate the 5804/5804 R/5810/5810 R centrifuge with the following rotors. Before use of
sample tubes, please note the manufacturer's specifications with regard to centrifugation
resistance (max. rcf).

2.4.1

Rotor A-4-81 (only 5810/5810 R)

Rotor A-4-81 with 500 mL rectangular bucket

Hint!

Eppendorf centrifuges may only be operated with rotors that are intended for use with the
corresponding centrifuge.

Only use rotors which bear the name of the centrifuge (e.g. 5804 R).

Max. g-force:

3,220 × g

Rotor A-4-81

Rectangular bucket
500 mL

Aerosol-tight cap

Max. speed:

4,000 rpm

Swing-bucket rotor
with 4 × 500 mL
rectangular buckets

Max. load per bucket
(adapter, tube and
contents):

780 g

Vessel

Vessel

Adapters

Adapter bottom
shape

Max. g-force

Capacity

Order no.

(international)

Tube diameter

Max. speed

Vessels per
adapter/rotor

Max. tube length
with/without
aerosol-tight
bucket cap

Centrifugation radius

Vessel

5810 745.004

flat

2,950 × g

1.5/2 mL

Ø 11 mm

4,000 rpm

20/80

43 mm/43 mm

16.5 cm

Blood collection tube

5810 746.000

flat

3,000 × g

1.2 to 5 mL

Ø 11 mm

4,000 rpm

20/80

108 mm/108 mm

16.8 cm

Vessel

5810 720.001

flat

3,000 × g

2.6 to 5 mL

Ø 13 mm

4,000 rpm

25/100

107 mm/108 mm

16.8 cm

Vessel

5810 747.007

flat

3,000 × g

2.6 to 7 mL

Ø 13 mm

4,000 rpm

18/72

108 mm/108 mm

16.8 cm

EN