5 operation – Eppendorf BioPhotometer User Manual

Page 22

Advertising
background image

64

5 Operation

Measuring procedure

Blank measurements remain stored until the date changes
(For further details, see "Measuring nucleic acids").

Standard measurements remain stored until they are overwritten
with new standard measurements. For the calculation of sample
measurements, the most-recently stored calibration is used.

In the following example, multiple-point calibration with 5 standards
in double determination and calibration calculation via non-linear
regression was programmed as a calculation procedure for the
Bradford method:

Measure blank

Measure
standards

Standard 1 /
first measurement

The first two lines of the display contain the standard that has just
been measured. The last two lines of the display contain the next
standard which is to be measured, with the nominal concentration.

Standard 1 /
second measurement

Blank

B R A D F O R D

B L A N K

0

.

0

0

0

A

Blank

Standard

or

Sample

Standard

B R A D F O R D

S T D 1 – 1

X X X X

µg/mL

N E X T:

S T D 1 – 2

X X X X

µg/mL

X

.

X

X

X

A

Standard

B R A D F O R D

S T D 1 – 2

X X X X

µg/mL

N E X T:

S T D 2 – 1

X X X X

µg/mL

X

.

X

X

X

A

05_Bedienung_e.fm Seite 64 Dienstag, 21. Februar 2006 10:09 Uhr

Advertising