2 performing the self-test – Eppendorf Self Test Function for Mastercycler nexus User Manual

Page 5

Advertising
background image

Self Test Function for Mastercycler® pro and Mastercycler® nexus

5

The status of the self-test function can be identified using the symbol in the navigation tree below

the cycler node.

Self-test function activated

Self-test function locked

The self-test function remains active until the self-test USB key is removed.

3.2 Performing the self-test

Requirement

A user is logged on (if the

PIN

function is activated).

The navigation tree is displayed.

The self-test function is unlocked (see p. 4).

1. Mount the thermoblock with an empty PCR plate and close the heated lid.
2. Open the cycler node.
3. Mark the

Self-test

node.

4. Click the

Start

softkey.

5. Click the

Start

softkey again.

The self-test runs automatically. Information on the currently running test is displayed.

CAUTION!

CAUTION! Risk of burns from hot thermoblock and heated lid.
The thermoblock and the heated lid very quickly reach temperatures above 95 °C.

Keep heated lid closed until the temperature of the thermoblock is below 30 °C.