4 application parameters, 1 temperature control, Application parameters 9.4.1 – Eppendorf ThermoMixer C User Manual
Page 43: Temperature control
Technical data
Eppendorf ThermoMixer
®
C
English (EN)
43
9.4
Application parameters
9.4.1
Temperature control
*Heating rate and cooling rate can be throttled (see Restricting ramp rates on p. 33).
Temperature control range
Minimum: 15 °C ±2 °C below ambient temperature
Maximum:100 °C
Temperature setting 1 °C – 100 °C, adjustable in steps of
1 °C
Temperature accuracy
Set temperature
20 °C – 45 °C
Set temperature
< 20 °C or > 45 °C
SmartBlock 1.5 mL
±0.5 °C
±1.0 °C
You can find information on the temperature accuracy of other SmartBlocks on the
website www.eppendorf.com.
Temperature homogeneity
In the range of 20 °C – 45 °C
max. ±0.5 °C for all positions of the thermoblock
Heating rate*
6.0 °C/min
The change of temperature in filled tubes is slower.
Cooling rate*
if the set temperature is
above ambient temperature
if the set temperature is
between ambient
temperature and 15 °C
below ambient temperature
2.5 °C/min
0.5 °C/min – 0.8 °C/min