Technical data

Eppendorf ThermoMixer

®

C

English (EN)

43

9.4

Application parameters

9.4.1

Temperature control

*Heating rate and cooling rate can be throttled (see Restricting ramp rates on p. 33).

Temperature control range

Minimum: 15 °C ±2 °C below ambient temperature
Maximum:100 °C
Temperature setting 1 °C – 100 °C, adjustable in steps of
1 °C

Temperature accuracy

Set temperature
20 °C – 45 °C

Set temperature
< 20 °C or > 45 °C

SmartBlock 1.5 mL

±0.5 °C

±1.0 °C

You can find information on the temperature accuracy of other SmartBlocks on the
website www.eppendorf.com.

Temperature homogeneity

In the range of 20 °C – 45 °C
max. ±0.5 °C for all positions of the thermoblock

Heating rate*

6.0 °C/min
The change of temperature in filled tubes is slower.

Cooling rate*

if the set temperature is
above ambient temperature

if the set temperature is
between ambient
temperature and 15 °C
below ambient temperature

2.5 °C/min

0.5 °C/min – 0.8 °C/min