Mixmate® — operating manual, Operating manual – Eppendorf MixMate User Manual

Page 10

Advertising
background image

MixMate® — Operating manual

10

WARNING!

WARNING! Injury from flying tubes and plates.
If the maximum permitted total weight of the mixing load is exceeded, plates or tubes

may become detached from the device.



Always ensure that tubes, plates and tube holder are well seated. The dimensions

of the plates used must comply with the ANSI/SBS standards for microplates



Mix the DWP and the tube holders 0.5 mL, 1.5/2.0 mL and PCR 96 with max. 2000

rpm.



Only mix mixing loads with a total weight of up to 80 g at maximum speed.



Only mix mixing loads with a total weight of 80 to 300g at speed of max. 2000 rpm.

WARNING!

WARNING! Injury from sample material being thrown out.
Sample material can be thrown out of open, improperly sealed or unstable tubes and

plates.



Only mix in closed tubes and closed plates.



Observe the nationally prescribed safety environment when working with

hazardous, toxic and pathogenic samples. Pay particular attention to the personal

protective equipment (gloves, clothing, goggles etc.), extraction, and the safety

class of the lab.

WARNING!

WARNING! Injury from improper vortex action.
Improper vortex action can destroy tubes or cause their content to be lost.



Only vortex intact and sealed tubes.



Never vortex tubes made of glass or other fragile material.

WARNING!

WARNING! Risk from incorrect supply voltage



Only connect the device to power supplies which correspond with the electrical

requirements on the nameplate.



Only use sockets with a protective earth (PE) conductor and a suitable mains cable.

NOTICE!

NOTICE! Damage to the display due to mechanical pressure.



Do not exert mechanical pressure on the display.

NOTICE!

NOTICE! Caution! Strong vibration.
When mixing at high speeds, items located near the device may be moved by the

vibrations of the work surface and, e.g., fall off the work table.



Do not place easily movable items near the mixer or secure them adequately.

NOTICE!

NOTICE! Damage to electronic components from spilled liquids.



Make sure that the vortex mat and the cover caps are fitted properly. If the vortex

mat is not fitted properly, contact your Eppendorf partner or the authorized

Technical Service.



If liquid has been spilt: Switch off the device, disconnect the power plug and

arrange for it to be cleaned by service personnel authorized by Eppendorf.