Technical data

9

Technical data

9.1

Specifications

9.1.1

Upright freezers specifications

* X = 0 for 120 V, 2 for 208 - 230 V, or 1 for 230 V
‡ Freezer set to -80 °C, ambient 20 - 25 °C at rated mains/electrical supply

Model No.

U410

U570

Part No.

U9260-000X*

U9270-000X*

Internal Dimensions:
Height x Width x Depth

1265 x 550 x 575 mm

49.8 x 21.6 x 22.6 in

1265 x 765 x 575 mm

49.8 x 30.1 x 22.6 in

External Dimensions:
Height x Width x Depth

1915 x 800 x 852 mm

75.3 x 31.5 x 33.5 in

1925 x 1025 x 852 mm

75.8 x 40.3 x 33.5 in

Capacity

410 Liters

14.5 cubic feet

570 Liters

20.0 cubic feet

Net Weight

235 kg

517 lb

265 kg

583 lb

Lock

Standard

Standard

No. Compartments

5

5

Interior

Stainless steel grade 304L

Alarms

High/Low temperature, power fail, battery low, filter clean, fault

Insulation Material

Urethane foam

Remote alarm port

Standard

Standard

RS-485 interface

Optional

Optional

Refrigerants:

High Stage Refrigerant: R404A / Low Stage Refrigerant: R508B

‡Power Consumption:

120 V electrical supply

540 Watts

590 Watts

208 - 230 V electrical
supply

540 Watts

590 Watts

230 V electrical supply

450 Watts

490 Watts

Mains/Power Source and Current Rating
120 V, 60 Hz

16.5 A

16.5 A

208 - 230 V, 60 Hz

8 A

9 A

230 V, 50 Hz

5 A

6 A

Pull Down Time: From +25°C to -85°C (freezer empty; 230 V, 50 Hz electrical supply)

5.3 hours

5.5 hours

Performance

-50 °C to -86 °C at 32 °C maximum ambient operating temperature

Environmental Conditions

All freezers use components tested to CE/UL specifications listed below:

Indoor use

Altitude up to 2000 m

Ambient temperature range 10 °C to 32 °C

Maximum relative humidity 80 % for temperatures up to 31 °C, decreasing linearly to
50 % relative humidity at 40 °C

Mains/power supply voltage fluctuations not to exceed ± 10 % of the nominal voltage

Installation category II

Pollution degree 2

9