E-25/E25R Classic Incubator Shakers

Operating Manual

Symptom(s)

Probable Cause(s) & Solution(s)

Load out of balance: unload all contents, then reload.

Operating noise

Loose component(s) in platform, subplatform and/or drive
assembly: call for service.
Heater fuse blown: replace.
Compressor fuse blown: replace.
Compressor over-pressure switch activated: call for service.
Ambient temperature too high or too low: cool or heat the room
as needed.
Defective heater: call for service.
Defective refrigeration system: call for service.

Incubator does not
reach set temperature.

Incorrect temperature indication (see below).
Defective RTD assembly: call for service.

Incorrect temperature
indication.

Defective main board: call for service.