Eppendorf SBS adapter User Manual

Sbs adapter — instructions for use

Advertising
background image

SBS adapter — Instructions for use

1

Instructions for use

This document supplements the instructions for use of the Centrifuge 5804/5804 R/5810/5810 R,

but does not replace them. Therefore, also read the instructions for use. The current version can

be found online at www.eppendorf.com.

Product description

The SBS adapter is designed for centrifugation of microplates and deepwell plates in SBS format

(127 mm x 85 mm) in the rotors A-2-MTP and A-2-DWP. The plates are centered in the carrier

using the adapter and secured against slipping.

Safety notes

Observe the maximum allowed loading height and load (see Technical data).

Load the rotor symmetrically. When doing so, follow the loading instructions included in the

instructions for use of the centrifuge.

Centrifugation

Place the SBS adapter in the plate carrier in so that the Eppendorf logo is legible.

Cleaning and Disinfection

Use a mild cleaning agent for cleaning.
Use alcohol (ethanol, isopropanol) or disinfectants containing alcohol for the disinfection.

Technical data

CAUTION!

CAUTION! Plate breakage due to incorrect positioning.
The plate frame may not protrude above the highest edges of the adapter on all sides.

Ensure that the plate is centered in the adapter.

Hint!

The SBS adapter cannot be autoclaved.

A-2-MTP rotor

A-2-DWP rotor

Max. g-force (rcf)/speed (rpm)

2,167 x g/4,680 rpm

2,227 x g/3,700 rpm

Max. centrifugation radius

88 mm

145 mm

Max. load (adapter, tube and contents)

380 g

380 g

Max. loading height

29 mm

89 mm

Material

Polyoxymethylene

Advertising