4.5.5

Settings Menu

Heating

This command is used to set the target temperature of the
instrument manually.

• Select or enter the Target temperature and click

Set and On

to

start instrument heating.

• Click the

Read

button to display the current temperature inside

the instrument or click the

Auto

check box to have it read

automatically.

• Click

Off

to stop heating.

• Click the down button

to display the heating graph and

click the up button

to hide it.

• Click the close button

to exit the Heating dialog box.

Properties

Select Properties to set a new alias name for the instrument.
Enter a new name in the New Alias field and click Set Alias to
confirm.

These settings take effect after restarting the software.

Filter Definitions

Measurement Mode

Select the appropriate filter position and enter the new wavelength,
bandwidth, and measurement mode for each new filter.
Choose from the dropdown list:

FI for fluorescence intensity,

ABS for absorbance measurements, and

Empty for filter-free positions.

Wavelength

Enter the filter wavelength.
For fluorescence intensity measurements, set the filter wavelength
within the allowed range of the instrument.
Absorbance filters are definable between 230 and 1000 nm
(Excitation only).

Bandwidth

Enter the bandwidth (nm) of the filter.

Description

This field can be used for individual user remarks about the filter,
e.g. filter name, application, etc.