Wiring the pg7, 20 ma option (loop powered), Battery replacement – APG PG7 user manual User Manual

Page 25

Advertising
background image

Rev. A, 12/09

PG7

25

Automation Products Group, Inc.

APG...Providing tailored solutions for measurement applications

Tel: 1/888/525-7300 • Fax: 1/435/753-7490 • www.apgsensors.com • [email protected]

Wiring the PG7

4-20 mA Option (loop powered):

NOTE 1: The supply voltage must be sufficient to maintain a minimum of 9

VDC after “dropping” voltage across the load resistance with the
output at 20mA. Example: If RL = 250 ohm then the “drop” is 0.02
Amps X 250 ohm = 5 volts. Therefore power supply minimum is 5 V +
9 V = 14V.

NOTE 2: Completion of the earth ground (Pin F) is recommended for proper

circuit protection.

A

F

B

E

C

D

6 Pin Connector

Battery Replacement:

Step 1: Press on the front bezel and turn counter-clockwise to release the bezel

from the gauge.

Step 2: Remove the front display to access the batteries.
Step 3: Replace the front display ensuring that the notch in the display aligns

with the tab on the gauge housing.

Step 4: Replace the bezel by lining up with the tabs on gauge while pressing

inward and twisting clockwise.