ClearPAGE Precast Gel Running Instructions 7-1-14

10

BUFFER FORMULATIONS

As an alternative to the buffers ClearPAGE sells, these formulations may be used to prepare buffers

yourself. Use high-quality, low-conductance ingredients. Do NOT use acid or base to adjust the pH!

Standard SDS Running Buffer, 20X for Reduced Samples (FB60500)

Ingredient

MW

Molarity

Qty/Liter

Tricine (free acid)

179.17

0.8M

143.4 g

Tris (free base)

121.14

1.2M

145.2 g

SDS (2%)

288.38

-

20.0 g

Sodium Meta-bisulfite

104.06

50mM

5.0 g

Ultrapure water (fill to)

-

-

1000ml

* pH should be between 8.4 and 8.5 at 25° C.

* For non-reduced samples (especially antibodies), omit the Sodium Meta-bisulfite

Turbo SDS Running Buffer, 20X FB13500

Ingredient

MW

Molarity

Qty/Liter

MPS (free acid)

209.26

0.6M

125.6 g

Tris (free base)

121.14

1.2M

145.2 g

SDS (2%)

288.38

-

20.0 g

Sodium Meta-bisulfite

104.06

50mM

5.0 g

Ultrapure water (fill to)

-

-

1000 ml

* pH should be between 8.3 and 8.4 at 25° C.

* For non-reduced samples (especially antibodies), omit the Sodium Meta-bisulfite

LDS Sample Buffer, 4x -FB31010

Ingredient

MW

Molarity

Qty/Liter

Glycerol (40%)

-

-

400 g

Ficoll-400 (4%)

-

-

40 g

Triethanol amine, pH7.6

149.2

0.8M

120.0

6 N HCL

36.46

-

93.0 g

Lithium Dodecyl Sulfate (4%)

-

-

40 g

EDTA Di-Sodium

372.2

2mM

7.44 g

Brilliant Blue G250 (0.025%)

-

-

0.25g

Phenol Red

-

-

0.25 g

Ultrapure water (fill to)

-

-

1000 ml

* pH should be between 7.7 and 7.8 at 25° C.

Advertising