Mega-Gel Instructions 12/11

7.

Apply gel solution to gel plate sandwich using a

syringe or pipette. If using a stacking gel, pour

desired height of running gel, then overlay a small

amount of dH

2

O or 0.1% SDS solution to top of

gel. After polymerization, rinse with buffer, add

stacking gel solution and insert comb. For regular,

single percentage gels, add polyacrylamide solution

to correct height, and insert comb. Allow gel to

polymerize, usually 20-30 minutes. Extra gel

solution in pipette or syringe can be monitored to

test polymerization of gel mix.

Leave the comb in the plate sandwich until just

before sample loading

8.

Disassembly. Hold the clamped plate assembly with

one hand. Remove the gasket by starting at one of the

top ends and pulling “up and out” on the gasket until

it releases from the plate, up to the bottom of each

of the white clamps. When each clamp is reached

DO NOT remove it, instead feed the gasket down

through the clamp body and repeat pulling up and out.

Continue feeding until the gasket is fully detached.

9.

Bottom spacer must be removed before electrophoresis. Using a

spatula, needle or extra spacer gently push down on angled part of bottom

spacer and remove. If gel is not to be used immediately, wrap entire plate

sandwich with plastic wrap tightly to seal and store at 4°C for up to a week.